UNI-PHASE 16, 48 and 128 (MAX) Systems

Automatic Systems
Using our 'exclusive' "Pulse-Phasic" Process!
- OPERATING INSTRUCTIONS -
LAST UPDATED 01/02/2007

INTRODUCTION
Dr. Marx, CSPSI

    Welcome to CS PRO. All of our UNI-PHASE Systems operate identically and produce equally high quality colloidal silver. Other than system size and electrode configuration, the only differences are batch volume and time. Our unique design and innovative process produces the highest quality CS, while providing the most consistent results of any other CS production method available, as well as being the fastest, most reliable and easiest to use. Our customers, many of whom have tried most of the other "generators" on the market, rate our systems and technical support as the best they have ever encountered - anywhere. This same high level of quality and service is available to each and every one of our customers. We are here if you need us and look forward to serving you.

NOTES
Please take a few moments to thoroughly read these Operating Instructions before using your ''UNI-PHASE System''

WARNING!! THIS IS A VERY POWERFUL SYSTEM! FORGETTING TO "STOP" THE PROCESS CAN RESULT IN DAMAGING THE TANK AND ELECTRODES. ALLOWING THE ELECTRODES TO "TOUCH" WITH POWER APPLIED CAN DAMAGE THE ELECTRONIC COMPONENTS. 'CS PRO' ASSUMES NO LIABILITY FOR DAMAGES DUE TO NOT FOLLOWING INSTRUCTIONS OR FROM NEGLIGENT OPERATION!

TIPS ON USING THE TDS-3 'TEST PROBE'

1. The Test Probe IS NOT WATERPROOF! Only the probe 'tip' is designed to be immersed in liquid.
2. DO NOT ATTEMPT TO CALIBRATE THE WATER TEST METER! The meter is pre calibrated.
3. Remove the protective cap from the bottom end of the probe.
4. Turn 'On' by pressing the ON-OFF button. It will display a "000" reading (A 5 PPM reading will appear as "005", etc.).
5. First, press the TEMP button and verify the water temperature is between 24-27deg. C (75-80 deg. F). If necessary, warm the batch water to at least 24 deg. C. Water 'slightly' warmer than 27 deg. C (up to 30 deg. C) is acceptable.
6. When using the meter, 'plunge' the tip of the probe vigorously into the liquid, 1/2 - 1 inch, 2-4 times, and stir. Then hold the meter still for the reading. The object is to ensure that the probe area is completely 'flushed' of any other liquid (previously tested) and only the liquid you are testing now is in the probe cavity.
7. Remember, NO TDS meter will test PPM silver! The readings you see are ONLY "conductivity", which you will be using in the calibration process. It has NOTHING to do with silver content! Over time the reading will actually drop considerably in the finished product. This is perfectly normal.
8. Turn the meter OFF when finished. If you forget, the meter will turn itself off automatically after a few minutes.
9. If you need to replace the batteries, GENTLY pry the top up and slide the battery assembly out of the meter body. You will see the two batteries. Replace with #357, 1.5V batteries available everywhere (WalMart, K-Mart, Radio Shack, etc.). Be sure to check the polarity of the old batteries and install the new ones the same way as the old ones were removed.
10. Your TDS-3 meter comes with a 3 year warranty. Be sure to keep the 'Users Guide' that comes with the meter so you will have the information necessary to obtain service. As we do not make the meters, the warranty is provided by the manufacturer. For the earlier 'blue' meters, call either HANNA Instruments Tech. Dept. at 800-426-6287 - X11 or X34 (warranty = 6 months), or Milwaukee Instruments at 252-443-3630 (warranty = 1 year), depending on which meter you have.

MAKING YOUR 'CALIBRATION SOLUTION'

1. Obtain a 'clean' container, with a wide mouth opening (could be a drinking glass) that will hold about 8 oz. for the UP-16 (or 16 oz. for the UP-48 and MAX 128). Some MAX 128 users prefer to just make an entire gallon. It will last indefinitely. You can use the system 'batch tank' for this step if you prefer; just be sure to rinse the tank thoroughly before proceeding to make a batch of CS.
2. Fill a container with 'steam distilled' water. Turn your test probe on, and test the water in the container (should be distilled quality water).
3. Begin adding the 'Calibration Powder' (ordinary 'Baking Soda' - we provide a small amount with each system to get you started), a VERY SMALL 'pinch' at a time. Stir thoroughly and take another reading. The objective is to build the meter reading to 800-850 PPM.
NOTE: Add VERY SMALL AMOUNTS, IN SEVERAL STEPS, as a very little goes a long way. If you exceed the target PPM, pour a little out and add small amounts of distilled water to achieve your 'target' reading of 800-850 PPM. Keep in mind, the 'quantity' of solution is not important and the reading does not need to be 'exact' as you are simply making a concentrate.
4. Fill the dropper bottle with the solution and put the rest into the bottle labeled "Calibration Solution" (for continued refilling of the dropper bottle).
5. When finished making the calibration solution, rinse the probe area of the meter in "tap water". This will remove most of the concentrate before using the meter for the 'lower' readings (calibrating the batch).
NOTE: Baking soda is found in virtually every household, is used in baking numerous foods, consumed, used as an antacid, mouth freshener, toothpaste, and it is totally harmless. It (sodium-hydrogencarbonate, aka. sodium-bicarbonate, aka. baking soda) does not react or combine with the silver in any way, unlike salt, aka. sodium chloride - which 'could' combine to form 'silver chloride' (one of the salts of silver). As it turns out, baking soda is capable of adding a tremendous amount of PPM/TDS, using a virtually insignificant amount of the powder!
INFO: Another very interesting side benefit of using 'baking soda' is that since distilled water will normally have a slightly acidic pH - generally 6.5 - 6.8 (7.0 is 'neutral'), any CS will remain slightly acidic if made without calibration, or using 'tap water' or 'salt' for the calibration. By using the baking soda method, the pH of a batch will be about 7.0-7.5 pH, well into the ideal 'alkaline area' which the human body prefers. In this case, the colloidal silver actually functions as a mild 'antacid'. We consider this to be a significant improvement in the production of CS.

'SETTING-UP' AND 'OPERATING' YOUR 'Uni-Phase' System

1. Take the ScotchBrite pad and (with a pair of scissors) cut it in half long ways, then across twice. This will give you 6 - 1"x1" pads, a perfect size for cleaning the electrodes.
2. Hold the "power head" over the sink with the electrodes angled downwards.  Wet one of the ScotchBrite squares and lightly 'swipe it down' the electrodes to remove discoloration. Check that they are reasonably straight, perpendicular, and parallel to each other. Wipe clean/dry with a paper towel.
3. Using a good quality 'Steam DISTILLED WATER' ("000 - 002 PPM" reading on the test probe), fill the batch tank. For the 16 and 48 this level will be just above the 'shoulder' (NOT up into the vertical 'neck'). For the MAX, fill to the fill line on the attached label (some DW bottles contain more than 1 Gal) being careful NOT to overfill. Generally, a well known 'Brand Name' of distilled water will provide the most consistent quality, even though you will occasionally find fluctuations in quality within the same brand. We suggest you try a number of different brand names. Eventually you will find the one that most consistently provides the quality of water you need. It is impossible to lower the conductivity of 'not good enough' water to an acceptable level. However, distilled water that is 'better' than necessary can always be adjusted 'up' during the calibration process! If the distilled water is not good enough to use in the process, keep for 'drinking', and try another bottle.
NOTE: The reason we want the initial reading of the water to be slightly lower than the final 'calibrated' reading is so that we can add at least a small amount of the calibration solution and slightly increase the pH.
NOTE: Temperature of the water will effect the operation of the system and the ultimate PPM/Ag. For best results, temperature should be between 75 to 80 degrees F (see step 5 of the TDS Probe Instructions, above).
4. Use the dropper to add the calibration solution to the batch water. For the MAX, you can add 10cc (using the dosage cup provided) of calibration solution and finish the calibration with the dropper. As you continue to 'add' calibration solution, 'stir' the batch water with the tip of the 'Test Probe' (use a spoon for the MAX), monitoring the reading. The final CALIBRATED READING should be 4-5 PPM.
NOTE: If you slightly "over calibrate" simply pour out about 1/3 of the water in the tank, refill with "new" water, and recheck the reading. If now slightly "under calibrated", re-add a small amount of calibration solution (a drop at a time) until it is correct. You may not need to do anything, depending on how far you went "over" on the first try. After you have gone through the calibration process a few times, you will get a "feel" for about how much of the calibration solution to add for achieving a given reading.
5. When 'calibration' is complete, install the 'power head' on the tank and place the system near the receptacle.
NOTE: These are the most versatile colloidal silver systems available You can plug the 'UNI-PHASE' systems (or the timer if you ordered the "Auto Version") into any standard 115VAC/60Hz. household receptacle, OR they can be used anywhere in the world (240, 230VAC - 50/60Hz.) by utilizing the 'Euro Timer Upgrade' and simply adapting the timer plug to fit your receptacle. You can also operate the UP-16 and 48 on 12VDC with an appropriate 'inverter'. This will permit using the system in a vehicle, or from any 12VDC source. You can also operate using the inverter with any 'solar panel' capable of putting out 20 Watts @ 15VDC. Because of the higher power requirement of the MAX, we do not recommend using it with an inverter without calling CS PRO first to determine the proper inverter capacity.
6. MANUAL OPERATION (see "Automatic" below) - Should your timer fail, you can use anything you wish to time the batch. The time does not have to be 'exact', but you should be as close as possible, give or take a few seconds (see "Time Table" below). Generally, running the system at 220/230 Volts will take approximately '2/3' the time required at 115 Volts. The "TABLE" (below) will give a 'General' guideline. As you become familiar with the operation, you will soon be able to 'visually' determine your selected batch 'stopping point'. If 'manually' operated, you MUST remember to physically unplug the system at the end of the appropriate running time.
7. If you have done everything correctly to this point, within the first fifteen seconds, you will notice one electrode (every other outer electrode with the MAX-128) begin to 'darken, and some very small bubbles forming on one of the other electrodes (the center electrode with the MAX-128). Shortly, you will begin to see bubbles streaming up the green electrode(s). Within another few minutes you will see very fine 'swirls' of particles beginning to form near the surface of the water and around the darkened electrode(s). As the process continues, (depending on the concentration you are running), you will see the clouds of particles increase. They will be slowly 'swirling' from one electrode to another, back and forth (see photo - appearance at approximately 15 PPM). This is the 'Pulse-Phasic' Process working. As the particles move from the 'phase field' of one electrode, and into the 'phase field' of another, they are being continuously reduced in size. SEE "OPERATING TIPS' - below.
8. When the Process has run the desired time, 'UNPLUG the UNI-PHASE System from the power source and bottle or store the CS you have just made. The shelf life should be 1-2 years! Clean the electrodes, wipe the tank dry with a soft paper towel, and you are finished.
UP-16
UP-16
UP-48
UP-48
MAX
MAX
Result
@115V
@230V
@115V
@230V
@115V
@230V
PPM Ag
20 min.
12 min.
25 min.
15 min.
11 min.
8 min.
10-11 (Recommended)
25 min.
15 min.
30 min.
18 min.
13 min.
9 min.
14-16
30 min.
18 min.
35 min.
21 min.
15 min.
10 min.
18-22
6A. AUTOMATIC OPERATION - (recommended)
All automatic systems now come with the 'Intermatic' Digital Programmable Start/Stop Tabletop Timer. Read and follow the manufacturers instructions. The EURO Rotary timer is self explanatory.

a. When the batch is calibrated and set up, PERFORM A FINAL VISUAL CHECK TO ENSURE THE ELECTRODES ARE STRAIGHT AND NOT TOUCHING!
b. For the US digital timer, set the 'start' and 'stop' time per the timer instructions (NOTE: Pay particular attention to AM and PM settings). NOTE: Allow a few minutes when setting the start time to be certain everything is ready. Then plug the system into the back side of the timer plug. It will start at the time you set and stop at the end of the cycle; whatever running time you allowed.
b1. For the EURO timer, plug the system into the receptacle on the back of the timer. Rotate the pointer quickly past (about 1/4 turn) and back to your desired running time and release. It will start immediately and turn off when reaching "0".
c. When the cycle is complete, either timer will turn off the power and your batch is done.
Always unplug the system from the timer before removing the powerhead from the tank.

MISCELLANEOUS OPERATING TIPS!

END OF BATCH READING: Although this reading has nothing to do with the Ag content of the batch, if you stir and test 'immediately' after finishing a 10-11 PPM/Ag batch, you should get a PPM/TDS reading of about 15-18. This is NOT PPM/Ag and this is ONLY a point of reference, enabling you to compare the end result of this batch to the last, the next, and so on. This reading is ONLY good at this point in time. Over time, the TDS reading on the stored product will gradually decrease. This has NOTHING to do with the actual silver content of your product! Once silver is added to the water, it remains the same, regardless of the TDS reading, which WILL diminish. This is due to the molecular restabilization of the water which was disrupted during the process and the subsequent cooling of the water!
PPM RECOMMENDATION: You can run the UNI-PHASE Systems for longer periods, and create higher PPM's, but we do not recommend ever going over 18 PPM. As you exceed 20 PPM, the size of the particles gradually increases, the 'bio-availability' goes down, a slight settling may begin, and shelf life is reduced. At this point, all you are doing is reducing the quality and effectiveness of your CS, and using excessive amounts of silver from the electrodes, drastically shortening their life. At the 99+% 'bio-availability' level of the CS produced by the 'Pulse-Phasic' Process, 10-12 PPM is more than enough to get the job done! In 15 years of experience and clinical application we have found no benefit to using concentrations over 10 PPM.
ELECTRODE CONDITIONING: If running for the recommended 10 PPM, the first batch you produce will be quite a bit stronger than normal (18-20 PPM/Ag) and darker, the second batch will be intermediate (13-15 PPM/Ag) and slightly darker than normal. By the third batch the electrodes will be conditioned and subsequent batches will be consistent. Do not discard the first and second batches. They are perfectly good to use. You cannot make a 'bad' batch of colloidal silver with these systems when following the operating instructions.
TEMPERATURE: Temperature is an important parameter in operating your system and you should always try to use water in the 75-80 deg. F (24-27 deg. C) range. This is often overlooked, particularly in the cooler months. Below 75F, performance will drop rapidly. Exceeding 80F slightly is OK as warmer has less effect than colder. Find a warm place in the house and allow the water time to equalize to this temperature. Stores are generally kept at 68F so the water will be quite cool. Allowing it to set in the warm area overnight will improve your results.
COLOR: PulsePhasic colloidal silver exhibits some unique characteristics. Up to 8-9 PPM, it will generally remain colorless. At about 10-12 PPM it may turn to very light yellow, while displaying a very light, almost indiscernible Tyndal (visible beam of a penlight shined through the liquid). Above 12 PPM, the yellow color may intensify. This is difficult to immediately ascertain since most all batches will finish colorless and the color will develop over time. Generally, the higher the PPM/Ag the more intense the yellow color becomes. Tyndal will also increase accordingly. The extreme "fineness" of the Tyndal indicates exceptionally small particles. Although batches may appear colorless at completion, the color may materialize slowly over a few hours. Color will only increase to a specific intensity and stop. This has no bearing or effect upon the quality or efficacy of the PulsePhasic CS.
ELECTRODE CARE: The electrodes should be gently "swiped" after each use. There will be slight discoloration on 2 of the electrodes (5 with the MAX-128), and a blackening of the 1 electrode (4 with the MAX-128), all easily removable, especially while wet. Do not scrub any harder than necessary to remove the discoloration. Excessive scrubbing only shortens the life of the electrodes. After every third or fourth batch, "remove" the electrodes for cleaning by pulling straight out of the sockets. DO NOT TWIST! To reinstall, push firmly into socket until they stop (about 1/4"). When removing used electrodes, make note of which 'end' came out of the socket and be sure the 'same end' is always reinserted. Do not flip electrodes "end for end", since as they wear and become thinner, they will no longer fit the socket. Whenever removing electrodes, to equalize the wear, reinstall them in different locations than removed from. The electrode that blackens wears the fastest. Make sure the electrodes are "straight" parallel and equidistant. This is particularly important with the longer electrodes of the UP-48 and MAX. Maintaining the correct distance between electrodes is important  to the quality of the product. NEVER ALLOW ELECTRODES TO TOUCH TOGETHER WHILE POWER IS APPLIED! This will create a "DEAD SHORT" and damage your system
SYSTEM CARE: All UniPhase systems require only minimal care. Aside from the electrodes, all that is needed to maintain the tank and powerhead are wiping with a paper towel before putting away. If necessary, the tank can be washed with a mild dish soap. Be sure to rinse thoroughly to avoid any soap residue. The powerhead may be periodically wiped clean with a damp cloth. NEVER IMMERSE THE POWERHEAD IN WATER! NEVER ALLOW WATER TO ENTER THE POWERHEAD THROUGH THE ELECTRODE SOCKETS!

!! - PRECAUTIONS - !!

THIS SYSTEM OPERATES AT POWER LEVELS WHICH ARE CAPABLE OF INFLICTING SERIOUS SHOCK, BURNS, FATAL INJURIES, ELECTROCUTION, AND POSSIBLE DEATH!

NEVER ATTEMPT TO REMOVE THE LID FROM THE TANK, OR TOUCH THE ELECTRODES, WHILE THE 'UNI-PHASE SYSTEM' IS PLUGGED IN! ALSO, NEVER PLUG-IN THE 'UNI-PHASE SYSTEM' UNTIL THE TANK IS PROPERLY SET-UP AND ADJUSTED, THE POWERHEAD IS IN PLACE AND YOU HAVE VISUALLY VERIFIED THAT NO ELECTRODES ARE TOUCHING EACH OTHER!

? - CHILDREN - ?

It is advised, that if you have small 'curious' children in the household, you should always remove the electrodes from the UNI-PHASE System before putting it away (preferably somewhere that they will not be able to find, or reach it), and keep the electrodes in a separate location.

NOTES!

The Colloidal Silver produced with the 'PULSE-PHASIC' Process, and the UNI-PHASE Systems, is many times better, more 'bio-available' at 10-15 PPM (100% vs. 10-20% for LVDC), and has a much longer (unstabilized) 'shelf life' than CS produced using the 'Low Voltage DC Process'. Colloidal silver produced with this process is equal in quality and efficacy to the '100% bio-availability' and 'indefinite shelf life' of the earlier CS PRO HVAC Process / Product.

STORAGE

Unlike LVDC CS, The 'Hybrid' CS produced by the 'Pulse-Phasic' Process can be stored in plastic bottles! There are bottles included in the UP-16 and 48 Packages for storage of your batch. It can also be kept for extended periods in the same 1 Gal. HDPE plastic bottles your DW comes in. For the MAX, since it makes a full gallon at a time, simply use the 1 gallon jug your DW came in. We have sample bottles which have been setting for over 7 years and are still in excellent (fresh) condition with no settling.

OBTAINING BOTTLES

We recommend: "E. D. Luce Pharmaceutical Packaging", <www.essentialsupplies.com> for bottles and caps of all kinds, plastic or glass. Their prices are very affordable, they have a reasonable 'minimum order', and reports are that they are very good to do business with.

WARRANTY & TECH SUPPORT

This CS PRO UNI-PHASE System is warranted against defects in materials and workmanship for a period of 1 (one) year from date of purchase. If operated in accordance with these instructions, a UNI-PHASE system will give you many years of satisfactory performance. If you have a problem, call CS PRO at 210-626-2546. If the problem cannot be solved over the telephone, we will issue you an 'RMA' number, authorizing return of the system for repairs or replacement, at the discretion of the manufacturer. No System will be accepted without an 'RMA' number.
'Products NOT manufactured by CS PRO' MUST be returned to their respective manufacturers for warranty service and repair.

CS PRO SYSTEMS, Inc.
2282 Old Pleasanton Rd.
San Antonio, TX 78264

Electrode Notes: Price of replacement electrodes will fluctuate depending on the market price of silver at the time of our order for inventory!
Current as of 12/31/06: UP-16 = $30 per set (3), UP-48 = $55 per set (3), MAX = $125.00 per set (9) - +$7 SH&I (in US) per order.
The UP-16 and UP-48 systems use the same 'Power Head' assembly. If uncertain as to which system you have when ordering electrodes,
the electrodes (3) for a UP-16 are about 3" long and for the UP-48 they are about 7" long. The MAX uses 9 - 7" electrodes.

Return to:
UniPhase 16&48 Page  -  UniPhase MAX Page

Last Updated - 01/02/2007
c) CS PRO SYSTEMS, Inc., 1996-2007